GC/MS/MS法结合SPME进样测定饮用水中亚硝胺类化合物

周懿, 尹戈, 孙谦, 范军, 邓晓丽, 黄涛宏. GC/MS/MS法结合SPME进样测定饮用水中亚硝胺类化合物[J]. 环境化学, 2022, 41(2): 770-772.
引用本文: 周懿, 尹戈, 孙谦, 范军, 邓晓丽, 黄涛宏. GC/MS/MS法结合SPME进样测定饮用水中亚硝胺类化合物[J]. 环境化学, 2022, 41(2): 770-772.
ZHOU Yi, YIN Ge, SUN Qian, FAN Jun, DENG Xiaoli, HUANG Taohong. Determination of nitrosamines in drinking water by GC/MS/MS combined with SPME injection[J]. Environmental Chemistry, 2022, 41(2): 770-772.
Citation: ZHOU Yi, YIN Ge, SUN Qian, FAN Jun, DENG Xiaoli, HUANG Taohong. Determination of nitrosamines in drinking water by GC/MS/MS combined with SPME injection[J]. Environmental Chemistry, 2022, 41(2): 770-772.

GC/MS/MS法结合SPME进样测定饮用水中亚硝胺类化合物

Determination of nitrosamines in drinking water by GC/MS/MS combined with SPME injection

  • 摘要: 本文利用岛津公司GCMS-TQ8050 NX三重四极杆气质联用仪以及AOC-6000自动进样器的SPME Arrow功能,建立了一种饮用水中亚硝胺类化合物的测定方法。水样经固相微萃取方式萃取后采用多反应监测模式(MRM)进行检测。11种亚硝胺类化合物在线性范围内线性关系良好,标准曲线相关系数(R)>0.999,检出限在0.55—79.58 ng·L−1。取校准曲线次低浓度点的标准品溶液连续进样6针,峰面积RSD均小于10 %。该方法简单方便,能够有效的监测生活饮用水中亚硝胺类物质的含量。
  • 加载中
  • 图 1  11种亚硝胺类化合物MRM图(浓度为2.0 ng·mL-1

    表 1  目标物的保留时间及MRM模式下优化的质谱参数

    编号组分名称CAS保留时间/min定量离子对定量离子CE/ V定性离子对定性离子CE/ V
    1N-亚硝基二甲基胺(NDMA)62-75-97.62074.00>44.10674.00>42.1021
    2N-亚硝基甲乙基胺(NEMA)10595-95-68.90088.00>71.10688.00>43.109
    3N-亚硝基二乙基胺(NDEA)55-18-59.833102.00>85.106102.00>44.1012
    4N-亚硝基二丙基胺 (NDPA)621-64-715.922130.00>113.106130.00>43.1015
    5N-亚硝基二丁基胺(NDBA)924-16-326.087116.00>99.106116.00>74.0012
    6N-亚硝基哌啶(NPiP)100-75-426.055114.00>84.109114.00>97.106
    7N-亚硝基-N-甲基苯胺(NMPhA)614-00-626.203106.00>77.0021106.00>51.1030
    8N-乙基-N-亚硝基苯胺(NEPhA)612-64-626.378106.00>77.1021121.00>106.1012
    9N-亚硝基吡咯烷(Npir)930-55-226.745100.00>55.109100.00>43.109
    10N-亚硝基吗啉(Nmor)59-89-227.652116.00>86.106116.00>56.1012
    11N-亚硝基二苯胺(NDPhA)86-30-632.966169.00>167.2024169.00>66.1024
    编号组分名称CAS保留时间/min定量离子对定量离子CE/ V定性离子对定性离子CE/ V
    1N-亚硝基二甲基胺(NDMA)62-75-97.62074.00>44.10674.00>42.1021
    2N-亚硝基甲乙基胺(NEMA)10595-95-68.90088.00>71.10688.00>43.109
    3N-亚硝基二乙基胺(NDEA)55-18-59.833102.00>85.106102.00>44.1012
    4N-亚硝基二丙基胺 (NDPA)621-64-715.922130.00>113.106130.00>43.1015
    5N-亚硝基二丁基胺(NDBA)924-16-326.087116.00>99.106116.00>74.0012
    6N-亚硝基哌啶(NPiP)100-75-426.055114.00>84.109114.00>97.106
    7N-亚硝基-N-甲基苯胺(NMPhA)614-00-626.203106.00>77.0021106.00>51.1030
    8N-乙基-N-亚硝基苯胺(NEPhA)612-64-626.378106.00>77.1021121.00>106.1012
    9N-亚硝基吡咯烷(Npir)930-55-226.745100.00>55.109100.00>43.109
    10N-亚硝基吗啉(Nmor)59-89-227.652116.00>86.106116.00>56.1012
    11N-亚硝基二苯胺(NDPhA)86-30-632.966169.00>167.2024169.00>66.1024
    下载: 导出CSV

    表 2  11种亚硝胺类化合物相关系数、检出限

    编号组分名称浓度系列/(ng·mL-1相关系数检出限/(ng·L-1
    1NDMA0.2、0.5、1、2、50.999424.67
    2NEMA0.1、0.2、0.5、1、20.999713.16
    3NDEA0.1、0.2、0.5、1、20.99977.50
    4NDPA0.1、0.2、0.5、1、20.99990.69
    5NDBA0.1、0.2、0.5、1、20.99940.55
    6NPiP0.1、0.2、0.5、1、20.99928.16
    7NMPhA0.1、0.2、0.5、1、20.999279.58
    8NEPhA0.1、0.2、0.5、1、20.999032.86
    9Npir0.5、1、2、5、100.999625.23
    10Nmor0.2、0.5、1、2、50.999114.86
    11NDPhA0.1、0.2、0.5、1、20.99910.95
    编号组分名称浓度系列/(ng·mL-1相关系数检出限/(ng·L-1
    1NDMA0.2、0.5、1、2、50.999424.67
    2NEMA0.1、0.2、0.5、1、20.999713.16
    3NDEA0.1、0.2、0.5、1、20.99977.50
    4NDPA0.1、0.2、0.5、1、20.99990.69
    5NDBA0.1、0.2、0.5、1、20.99940.55
    6NPiP0.1、0.2、0.5、1、20.99928.16
    7NMPhA0.1、0.2、0.5、1、20.999279.58
    8NEPhA0.1、0.2、0.5、1、20.999032.86
    9Npir0.5、1、2、5、100.999625.23
    10Nmor0.2、0.5、1、2、50.999114.86
    11NDPhA0.1、0.2、0.5、1、20.99910.95
    下载: 导出CSV

    表 3  11种亚硝胺类化合物的重复性结果

    编号.组分名称峰面积
    123456RSD/%
    1NDMA1161511524115011257811678126924.61
    2NEMA2339324736232722383822266215704.86
    3NDEA6377460473607616497162374609582.94
    4NDPA2828783015592837133062672841452692624.73
    5NDBA5211995499985303435643545277765060953.91
    6NPiP3141633999336503315628545298826.97
    7NMPhA7794358409968176038806538007927652405.19
    8NEPhA1220366129689512217261322022122902711655634.60
    9Npir5939671056526722534856099.80
    10Nmor1055111727114351077810698118445.05
    12NDPhA8194248600168245079061548438447933044.64
    编号.组分名称峰面积
    123456RSD/%
    1NDMA1161511524115011257811678126924.61
    2NEMA2339324736232722383822266215704.86
    3NDEA6377460473607616497162374609582.94
    4NDPA2828783015592837133062672841452692624.73
    5NDBA5211995499985303435643545277765060953.91
    6NPiP3141633999336503315628545298826.97
    7NMPhA7794358409968176038806538007927652405.19
    8NEPhA1220366129689512217261322022122902711655634.60
    9Npir5939671056526722534856099.80
    10Nmor1055111727114351077810698118445.05
    12NDPhA8194248600168245079061548438447933044.64
    下载: 导出CSV
  • [1] 汤庆会, 余沛芝, 陆坤, 等. 固相萃取-气相色谱串联质谱测定饮用水中N-亚硝胺类消毒副产物[J]. 净水技术, 2018, 37(10): 28-32

    TANG Q H, YU P Z, LU K, et al. Detection of N-nitrosamine disinfeedon by-products in dnnking warer by SPE-GC/MS[J]. Water Purification Technology. 2018, 37(10): 28-32.

    [2] NAWROCKI J, ANDRZEJEWSKI P. Nitrosamines and water[ J]. Journal of Hazardous Materials, 2011, 189(1-2): 1 -18.
    [3] BEI E, SHU Y, LI S, et al. Occurrence of nitrosamines and their precursors in drinking water systems around mainland China. Water Research, 2016, 50(1) : 168 - 175.
    [4] 科技. 中国消化道癌症高发或与喝水有关 [J]. 科学大观园, 2016(21): 6. doi: 10.3969/j.issn.1003-1871.2016.21.003

    Technology. High incidence of gastrointestinal cancer in China may be related to drinking water [J]. Grand Garden of Science, 2016(21): 6(in Chinese). doi: 10.3969/j.issn.1003-1871.2016.21.003

    [5] RICHARDSON S D. Water analysis: Emerging contaminants and current issues [J]. Analytical Chemistry, 2009, 81(12): 4645-4677. doi: 10.1021/ac9008012
    [6] TRAIF K, WEILAND S, BUNGERS M, et al. Exposure to hish concentrations of nitrosamines and cancer mortality among a cohort of rubber workers [J]. Occupation and Environmental Medicine, 2000, 57(3): 180-187. doi: 10.1136/oem.57.3.180
    [7] 张立峰. 咸鱼致癌, 都是亚硝胺惹的祸 [J]. 抗癌之窗, 2017(6): 32-34.

    ZHANG L F. Salted fish cause cancer. It's all caused by nitrosamines [J]. Cancer Frontier, 2017(6): 32-34(in Chinese).

  • 加载中
图( 1) 表( 3)
计量
  • 文章访问数:  1747
  • HTML全文浏览数:  1747
  • PDF下载数:  66
  • 施引文献:  0
出版历程
  • 刊出日期:  2022-02-27
周懿, 尹戈, 孙谦, 范军, 邓晓丽, 黄涛宏. GC/MS/MS法结合SPME进样测定饮用水中亚硝胺类化合物[J]. 环境化学, 2022, 41(2): 770-772.
引用本文: 周懿, 尹戈, 孙谦, 范军, 邓晓丽, 黄涛宏. GC/MS/MS法结合SPME进样测定饮用水中亚硝胺类化合物[J]. 环境化学, 2022, 41(2): 770-772.
ZHOU Yi, YIN Ge, SUN Qian, FAN Jun, DENG Xiaoli, HUANG Taohong. Determination of nitrosamines in drinking water by GC/MS/MS combined with SPME injection[J]. Environmental Chemistry, 2022, 41(2): 770-772.
Citation: ZHOU Yi, YIN Ge, SUN Qian, FAN Jun, DENG Xiaoli, HUANG Taohong. Determination of nitrosamines in drinking water by GC/MS/MS combined with SPME injection[J]. Environmental Chemistry, 2022, 41(2): 770-772.

GC/MS/MS法结合SPME进样测定饮用水中亚硝胺类化合物

  • 岛津企业管理(中国)有限公司,上海,200233

摘要: 本文利用岛津公司GCMS-TQ8050 NX三重四极杆气质联用仪以及AOC-6000自动进样器的SPME Arrow功能,建立了一种饮用水中亚硝胺类化合物的测定方法。水样经固相微萃取方式萃取后采用多反应监测模式(MRM)进行检测。11种亚硝胺类化合物在线性范围内线性关系良好,标准曲线相关系数(R)>0.999,检出限在0.55—79.58 ng·L−1。取校准曲线次低浓度点的标准品溶液连续进样6针,峰面积RSD均小于10 %。该方法简单方便,能够有效的监测生活饮用水中亚硝胺类物质的含量。

English Abstract

  • 亚硝胺类物质经常在许多国家的饮用水氯胺消毒过程中被检测到[1]。自来水厂的消毒过程常常使用含氯消毒剂,在氯的作用下水中有机氮化物会变成亚硝胺类的消毒副产物[2-3]。另一方面,大量的工业废水和生活污水使得饮用水水源污染加重。有检测显示,原水中就已出现较高浓度的有机氮,导致出厂水亚硝胺浓度的升高[4]

    亚硝胺类化合物由于其能够与DNA反应使其烷基化,进而导致癌症的发生,被国际癌症研究中心判定为2A类致癌物[5-6]。另有很多流行病学研究表明,亚硝胺类化合物与消化系统癌症有较明显的相关性[7]

    水中的亚硝胺类消毒副产物通常含量很低,并且具有亲水性,难以从复杂的水体环境中分离出来,因此,亚硝胺类物质通常不能直接进行仪器检测,需要进行样品预处理。一般采用液液萃取、固相萃取等前处理方法对1 L水样进行提取富集,操作过程复杂,耗费时间长,灵敏度低。

    针对上述情况,本文采用岛津AOC-6000+GCMS-TQ8050 NX,建立一种分析饮用水中亚硝胺类化合物的检测方法,该方法简单方便,绿色环保,抗干扰能力强,检出限低,可为水质分析中亚硝胺含量的测定提供借鉴。

    • 三重四极杆气质联用仪:GCMS-TQ8050 NX;自动前处理装置:AOC-6000。

      SPME:萃取纤维:20 mm/120 μm DVB/CarbonWR/PDMS;萃取温度30℃;孵化时间5 min; 萃取平衡时间50 min;震荡速率500 r·min−1 ;进样口解析温度250℃ ;进样口解析时间2 min。

      GC/MS/MS分析条件:进样口温度250℃; 载气控制方式:恒线速度(36.1 cm·s−1);进样方式:分流进样,分流比5∶1;色谱柱:InertCap Pure Wax,(30 m×0.25 mm×0.25 μm)(SGLC公司);柱温程序:40℃_10℃·min−1 _80℃_1℃·min−1 _100℃_20℃·min−1 _240℃(5 min);离子源温度230℃ ;接口温度250℃ ;采集方式:MRM,离子信息见表1

    • 准确移取10 mL 水样于20 mL 顶空瓶中,压盖,平衡5 min后,在30 ℃下高速搅拌萃取50 min,进样分析。

    • 标准色谱图如图1所示,化合物相关信息见表1.

    • 用超纯水为溶剂配制不同浓度的混合标准工作溶液,各组分浓度如下表2所示。按优化后的固相微萃取条件和仪器条件进行萃取分析。以浓度为横坐标,定量离子峰面积为纵坐标进行线性拟合,部分组分标准曲线如下所示。以3倍信噪比(S/N)计算各目标物的检出限,各物质标准曲线浓度梯度、检出限和标准曲线相关系数如下表2所示.

    • 取校准曲线浓度系列的次低浓度标准溶液,重复进样6次,考察仪器重复性,各组分峰面积及RSD见表3

    • 取某地区的饮用水,采用以上方法进行11种亚硝胺类化合物的检测,得到样品色谱图,均未检测出目标化合物。

    • 本方法采用岛津AOC-6000+GCMS-TQ8050 NX检测饮用水中亚硝胺类物质,在0.1—10 μg·L−1浓度范围内标准曲线线性良好,相关系数均在0.999以上,检出限在0.55—79.58 ng·L−1。取校准曲线浓度系列的次低浓度标准溶液连续进样6针,峰面积RSD均小于10 %,精密度良好。该方法简单方便,无需使用有机溶剂,绿色环保,能够有效的检测饮用水中亚硝胺类物质的含量。